C-CO2 Colorimetric Carbon Dioxide detector now available as an optional drop in with every adult resuscitator bag. Breath to breath color change reflects ventilation of the
patient and is an adjunct to clinical assessment when verifying correct endotracheal tube placement.
Versatile reservoir locations permit you to choose the resuscitation style of greatest comfort.
Valve system delivers high FIO2’s during all modes of operation (manual ventilation and spontaneous breathing).
Swivel patient connector reduces tracheal tube torque during handoff or transport.
Product Specifications:
Product Description
Adult Self-Inflating Manual Resuscitators
Ordering Information (10/Case)
Tail Reservoir Hose--7500 Series
Tail Reservoir Bag--7500K Series
Patient Size
>30 kg
Ventilatory Frequency
37-43 BPM at 600 ml VT
Stroke Volume Range
900 to 1100 ml (one hand); 1000-1400 ml (two hands)
Resuscitator Bag Volume
1750 ml
Insp. Pressure Limit
Available with optional Safety Relief Valve that "pops off" at 35-45 cm H2O at 15 LPM when valve is open
Delivered Oxygen
.98 +/- .02 at 15 LPM, 12 BPM, 600 ml VT, controlled or spontaneous ventilation
Dead Space
Less than 7.5 ml
Expiratory Resistance
2.5 +/- 1 cm H2O at 50 LPM (without accessories)
Inspiratory Resistance
4.0 +/- 1 cm H2O at 50 LPM
Operating Temperature
0 - 123oF
Storage Temperature
-40oF to 140oF
Connection Sizes
Patient: 15mm/22mm
Oxygen: DISS fitting
Expiratory Port: 30 mm OD
Optional Pressure Monitoring Port: 4 mm OD
Vital Signs is renowned as a product innovator; and we deliver to our customers single-patient use products of unsurpassed quality. With more than 30 years of leadership, Vital Signs continues our commitment to understanding our customers' needs and providing value-based solutions.